|
Recent CasesSupreme Court of the State of Washington City of Seattle v. Fawn Allison FACTS The trial court suppressed the breath test results. The city appealed and this court granted direct review. DISCUSSION When the DataMaster replaced the Breathalyzer machine, a similar set of foundational requirements were adopted, ultimately codified in chapter 448- 12 WAC , the precursor to the current WAC regulations in chapter 448-13 WAC. The new machine, and its supporting protocols, was challenged and upheld in State v. Ford, 110 Wn.2d 827, 755 P.2d 806 (1988) and Straka, 116 Wn.2d 859. The State Toxicologist has the delegated authority to designate proper methods for performing the analysis of a person's blood or breath. RCW 46.61.506(3). In Ford, this court observed that 'the Legislature has mandated that the analysis of breath or blood is valid if it is performed 'according to methods approved by the state toxicologist'.' Ford, 110 Wn.2d at 833 (quoting RCW 46.61.506(3)). '{W}hen the protocols . . . and existing Code provisions are followed, there is sufficient assurance of accuracy and reliability of the test results to allow for general admissibility of test results.' Straka, 116 Wn.2d at 870. The relevant procedures are found in several WAC provisions relating to the use and proper functioning of the DataMaster equipment. Former WAC 448-13- 040 provides in part:
The issue is whether the breath test documents printed by the DataMaster in these cases, which read 'SIM TEMP 34c +/- .2c: YES', are sufficient to establish a foundation for admissibility of the test results under former WAC 448-13-040. The defendants' primary argument is that use of the word 'be' in the WAC requires the city to prove the actual temperature of the simulator solution. The city argues that the WAC requires the operator only to determine whether the temperature reading on the DataMaster machine is within the prescribed range and that evidence of thermometer variances bears upon the reliability of the test results in a particular case, and goes to weight, not admissibility of the evidence. Accordingly, the city argues that the test results were improperly suppressed.
Read in context, the language at issue in former WAC 448-13-040 is a directive to the DataMaster operator to observe whether the temperature of the simulator solution is 34 degrees centigrade plus or minus 0.2 degrees centigrade. If the thermometer registers a temperature within that range, the operator may obtain a subject's breath sample. The rule does not require the operator to ascertain the actual temperature of the simulator solution. This makes sense since the DataMaster does not record actual temperature. Rather, when the operator observes a temperature within the prescribed range, he or she may proceed with the test when the machine records 'SIM TEMP 34c +/- .2c: YES' on the breath test document. Contrary to defendants' position, the rule does not require, nor provide a method for, determining the actual temperature or the accuracy of the thermometer. Nor does the actual operation of the machine lend itself to such a regulation. If we accepted the defendants' interpretation of former WAC 448-13-040 the breath test results would be inadmissible in every case, unless, as the defendants suggest, an operator additionally records the actual temperature reading and then, using the thermometer variance of +/- 0.1, calculates whether the actual temperature falls within the rule's range. We do not believe that the toxicologist intended to promulgate a rule containing requirements that are beyond the performance abilities of the machine or that require an operator to take steps that are not included in the rule. Accordingly, we conclude that compliance with former WAC 448-13-040 requires only that the breath test ticket reflect that the temperature was within the specified range. This construction is consistent with the language and purpose of the rule, which is to set forth the steps that must be followed to obtain a valid test, as well as with the actual operation of the machine. The defendants argue, though, that 2001 amendments to WAC 448-13-040 support a conclusion that the toxicologist must have intended that the simulator solution temperature be exactly 34 degrees centigrade plus or minus 0.2 degrees centigrade to produce a valid test result under the former regulation. In an emergency rule making order amending WAC 448-13- 040, the State Toxicologist specifically changed the temperature range for the simulator solution to 34 degrees centigrade plus or minus 0.3 degrees to account for the 0.1 degree variance built in to the thermometers.1 In issuing the order, the State Toxicologist wrote:
In light of the ruling from Renton, the acceptable variance specified in the WAC for the temperature of the simulator at the time of the test is being expanded to be plus or minus 0.3 degrees centigrade. This permits the continued use of the current instrument standard of plus or minus 0.2 degrees centigrade at the time of the test, and recognizes the limits of accuracy of the thermometer. An additional provision requires the thermometers to be certified at least annually. These provisions are intended to address the concerns of the courts on the admissibility of breath alcohol test results, and to permit all future tests to be considered as evidence provided they meet these and the existing requirements of the administrative code. Proposed Rules to WAC 448-13-040, St. Reg. 01-11-134
(May 23, 2001). In response to two comments received regarding the adoption
of the new rule, the State Toxicologist explained further: Response: This question reflects a misunderstanding of the issue ruled on by judges in Renton and Bellevue district courts in King County. These courts found that there was no explicit recognition in the WAC of the limits of accuracy of the mercury in glass thermometers used in the breath test equipment, and that thus, any variability in the thermometer must be included within the acceptable range. By enlarging the acceptable range to 0.3 degrees centigrade, but continuing to require a thermometer reading of 34 degrees plus or minus 0.2 degrees in the operation of the program, the potential variance in the thermometer is explicitly recognized and allowed for. Changing this allowable range does not introduce any additional inaccuracy into the measurement of the subject's breath alcohol concentration.2 Id. at ex. 5 (Permanent Rule Adoption for WAC 448-13-040, St. Reg. 01-17- 009 (Aug. 2, 2001)). The language of the amendment and the order do not support the defendants' construction. As the toxicologist's order states, his decision to amend the rule to account for the variation in thermometers was based on the decisions of two courts that any variation had to be included within the rules. His expressed concern was that other courts might make similar rulings. As his response to questions makes clear, the amendment changing the temperature range to expressly recognize thermometer variances does not change the requirement that the temperature fall within the range of 34 degrees centigrade plus or minus 0.2 degrees centigrade. The amendment does not undercut reliability of the test under the former rule. Finally, the defendants' construction is not necessary to ensure accuracy or reliability for the purpose of general admissibility. In Baker the officers waited 14 minutes before administering a breath test. This court held that the foundation for admissibility was not met because the WAC requires a 15 minute waiting period and literature from experts in the field established that the 15 minute waiting period is necessary to give an accurate measure of the concentration of alcohol in the blood. Baker, 56 Wn.2d at 856. Similarly, in Ford, 110 Wn.2d at 833, and Straka, 116 Wn.2d at 870-71, this court emphasized that the purpose of requiring compliance with the regulations is to assure accuracy and reliability of the test results for general admissibility. In this case, the 2001 amendment to WAC 448-13-040 and the accompanying emergency order indicate that the toxicologist did not believe minor variations in the machine's thermometers affect the accuracy or reliability of the test results. Based on the language of former WAC 448-13-040, we hold that the breath test documents in these cases constitute a sufficient foundation for admissibility of the test results and arguments as to reliability of the particular test results are questions for the jury. Straka, 116 Wn.2d at 875; see State v. Watson, 51 Wn. App. 947, 951, 756 P.2d 177 (1988) ('once the {breath test} evidence is rendered admissible by demonstrating compliance with administrative regulations, the defendant may attack the machine's reliability by presenting evidence to the jury') (citing State v. Franco, 96 Wn.2d 816, 639 P.2d 1320 (1982); City of Bremerton v. Osborne, 66 Wn.2d 281, 401 P.2d 973 (1965)). The trial court's order suppressing breath
test results is reversed and the cases are remanded for further proceedings. ........................ 1 The amended WAC 448-13-040 now requires: Prior to the start of the test the operator must verify that the thermometer, certified per WAC 448-13-035, indicates that the temperature of the simulator solution is thirty-four degrees centigrade plus or minus 0.3 degrees centigrade. 2 The emergency order postdated the tests in this case and neither party argues for its application to these cases. |
|||||||||||||||||||||||||||||||
© 2006 The Law Offices of Kevin Trombold, LLC. |
||||||||||||||||||||||||||||||||